Medical wireless foot switch
MKF SW2.4LE-MED SK13
Developed for the actuation of surgical or cosmetic laser devices to IEC 60601-2-22
Latchable protective guard prevents unintended actuation of the control unit
Wireless signal transmission per steute wireless low energy
Power supply via 3 commercially available batteries (AA)
Typical battery life: 230 days
Battery compartment can be opened by hand
Photo shows optional accessories
Configuration options
Technical data
General technical data
Applied standards
IEC 60601-1, ANSI/AAMI ES60601-1, IEC 62304, IEC 60529
Enclosure
thermoplastic, impact resistant, self-extinguishing UL 94 V-0, Pantone Cool Gray 1C
Pedal
thermoplastic, impact resistant, self-extinguishing UL 94 V-0/V-2
Antenna
internal antenna
Degree of protection
IPX8 (IEC/EN 60529)
Voltage source
3 commercially available batteries, type AA
Switching elements
reed contact / micro switch / Hall sensor
Switching system
1 – 2 NO contacts
Mechanical life
> 1 million operations
Conformity
Global: cCSAus IEC 60601-1, -1-2, -2-22, -2-43
Europe: CE EN 300 328; EN 62479; EN 301 489-1; EN 62368-1
USA: FCC Title 47 CFR, Part 15
Canada: IC RSS-247, Issue 1
Japan: ARIB STD-T66
Wireless technology
Frequency
2.4 - 2.4835 GHz
Channel bandwidth
2 MHz
Wireless range
10 m (typical)
Output power
3 dBm - 7 dBm
Input sensitivity
-93 dBm
HF data rate
1 Mbps
Modulation principle
GFSK, adaptive frequency hopping of 40 channels
Interface data rate
115.2 kBd (UART)
Note
Since this article is part of a medical product, it can only be validated in combination with the overall system. Proof of conformity to the Medical Device Regulation (MDR 2017/745) can therefore only be handled by the customer marketing the system. At the request of the customer, steute can affix a CE mark with the 4-digit number of the appropriate notified body.