REC SW2.4LE-MED 5VDC PCB

  • Wireless receiver for medical applications
  • Circuit board for integration in medical device
  • 1 digital validate output (relay)
  • 1 serial interface (RS 232)
Technical data
Applied standards
IEC 60601-1; ANSI/AAMI ES60601-1; IEC 62304; IEC 60529
Mechanical life
> 1 million operations
Rated operating voltage Ue
5 VDC
Rated operating current Ie
not activated < 100 mA; activated < 200 mA
Rated operating power Pe
500 mW
Outputs
1 potential-free NO contact: K1 Reed-Relay
Switching capacity
max. 10 W
Switching voltage
24 VAC/DC
Switching current
max. 0.5 A
Antenna
external antenna
Wireless range
10 m (typical)
Radio technology
Frequency
2.4 - 2.4835 GHz
Channel bandwidth
2 MHz
Modulation principle
GFSK, adaptive frequency hopping of 40 channels
HF data rate
1 Mbps
Interface data rate
115.2 kBd (UART)
Output power
3 dBm - 7 dBm
Input sensitivity
-93 dBm
Conformity
Global: cCSAus IEC 60601-1, -1-2, -2-22, -2-43
Europe: CE EN 300 328; EN 62479; EN 301 489-1; EN 62368-1
USA: FCC Title 47 CFR, Part 15
Canada: IC RSS-247, Issue 1
Japan: ARIB STD-T66
Note

Since this article is part of a medical product, it can only be validated in combination with the overall system. Proof of conformity to the Medical Device Regulation (MDR 2017/745) can therefore only be handled by the customer marketing the system. At the request of the customer, steute can affix a CE mark with the 4-digit number of the appropriate notified body.

Get in touch?
Please contact us!

Guido Becker
Division Manager Meditec

+49 5731 745-457
g.becker@steute-meditec.com