MKF-MED SK12

The special highlight of the variant MKF-MED SK12 with movable protective shield is  the  low  height  and  the  protection  against  accidental  actuation  in closed position.A quick actuation of the protective shield is sufficient to enable the operating  mode.  This  variant  is  suitable  especially  for  transportable laser  units.

  • Robust fibreglass-reinforced plastic foot control
  • GRP protective flap provides maximum protection from unintended actuation (IEC 60601-2-22)
  • Minimal design
  • Waterproof: IPX8 (IEC 60529)
  • Digital or analogue output signals available
  • Different RAL colours available

Individually configurable

Technical data

Applied standards
IEC 60601-1; ANSI/AAMI ES60601-1; IEC 60529
Pedal
shock-proof thermoplastic, UL 94 V-0/V-2
Protective flap
shock-proof thermoplastic, UL 94 V-0, RAL 7035
Housing
GD-Zn alloy, RAL 7035
Degree of protection
IP X8 to IEC/EN 60529
Switching elements
reed contact / micro switch / Hall sensor
Switching system
1 – 2 NO contacts / 1 change-over / analogue output 0 … 10 VDC,
0 … 20 mA, 4 … 20 mA
Mechanical life
> 1 million operations
Connection
2 m cable, other lengths optionally available
Switching voltage
max. 25 VAC / 60 VDC
Switch-on current
max. 1 A
Switching capacity
Reed contacts: max. 20 VA, micro switch: max. 30 VA
Supply voltage Hall sensor
15 … 30 VDC / max. 25 mA
 
 